ISO 15189:2022 Checklist for Medical Labs (Clause by Clause)

A clause-by-clause ISO 15189:2022 checklist for medical labs: what assessors look for, the evidence to have ready, top audit findings and changes from 2012.

An ISO 15189:2022 checklist works through clauses 4 to 8 of the standard: general requirements, structure and governance, resources, processes and the management system. For each requirement you confirm what assessors will test and which objective evidence you can show on the day.

Evidence is where most labs struggle. Findings rarely arise because a policy is missing; they arise because the record proving the policy works is incomplete, out of date or contradicts what staff say at the bench.

Use the tables below as a gap analysis a few months before your assessment, then again a few weeks out.

What changed from ISO 15189:2012

ISO 15189:2022 was published in December 2022 and replaced the 2012 edition. The three-year transition period ended in December 2025; confirm your status with your accreditation body. The main changes are:

  • Structure aligned with ISO/IEC 17025:2017, with management system requirements moved to clause 8.
  • Explicit risk management (5.6 and 8.5), focused on risks to patients across the whole examination process.
  • Stronger patient focus, including a new clause 4.3 on requirements regarding patients.
  • Point-of-care testing (POCT) incorporated: requirements formerly in ISO 22870 are now covered, with additional POCT requirements in Annex A.
  • Less prescriptive about named procedures, with more emphasis on outcomes.

Clause 4: General requirements

ClauseWhat auditors look forEvidence to have ready
4.1 ImpartialityThreats to impartiality identified and managed, including commercial and financial pressuresImpartiality risk assessment, declarations of interest, management commitment
4.2 ConfidentialityPatient information protected and released only to authorised partiesConfidentiality agreements, LIS access controls, release-of-information records
4.3 Requirements regarding patientsPatient welfare, information about the examination process, disclosure of incidents that could affect patientsPatient information, consent practice, incident disclosure records

Clause 5: Structural and governance requirements

ClauseWhat auditors look forEvidence to have ready
5.1 Legal entityA legally responsible entityLicence, registration, organisation chart
5.2 Laboratory directorQualified director with defined responsibilities and documented delegationDirector CV, job description, delegation records
5.3 Laboratory activitiesDefined scope, including advisory and interpretive servicesScope of examinations, advisory records
5.4 Structure and authorityClear roles, reporting lines and deputiesOrganogram, job descriptions
5.5 Objectives and policiesMeasurable objectives that reflect user needsObjectives with targets, quality indicator data
5.6 Risk managementRisks to patients identified and mitigatedRisk assessments for pre-, intra- and post-examination steps

Clause 6: Resource requirements

ClauseWhat auditors look forEvidence to have ready
6.2 PersonnelCompetence criteria, training, authorisation, competence assessment, continuing educationTraining files, competence assessments, authorisation matrix
6.3 Facilities and environmental conditionsSuitable, safe facilities; monitoring where conditions affect resultsTemperature and humidity logs, cleaning records
6.4 EquipmentAcceptance testing, maintenance, handling of defective equipmentEquipment inventory, acceptance and maintenance records
6.5 Calibration and metrological traceabilityCalibration programme with traceability to higher-order references where availableCalibration certificates, calibrator traceability statements
6.6 Reagents and consumablesReceipt, storage, acceptance testing, lot-to-lot checksLot acceptance records, storage logs, inventory
6.7 Service agreementsAgreements with users, and with POCT operators where relevant, reviewedAgreements and review records
6.8 Externally provided products and servicesSuppliers, referral labs and consultants evaluated and monitoredApproved supplier list, referral lab evaluations

Clause 7 is where assessors spend most of their time, usually by tracing a sample from request to report.

ClauseWhat auditors look forEvidence to have ready
7.2 Pre-examinationRequest information, patient preparation, collection, transport, acceptance and rejection criteriaCollection manual, rejection log, transport records
7.3.2 and 7.3.3 Verification and validationMethods verified before use; validation for non-standard or modified methodsVerification and validation reports
7.3.4 Measurement uncertaintyMU evaluated for quantitative examinations and reviewedMU calculations with review dates
7.3.5 Biological reference intervals and clinical decision limitsIntervals appropriate to the population servedSource or verification data, review records
7.3.7 Ensuring validity of resultsIQC able to detect clinically significant error, EQA participation, comparability between systemsIQC charts and rules, EQA investigations, comparability studies
7.4 Post-examinationAuthorised review and release; prompt communication of critical resultsCritical result log, amended report records
7.5 Nonconforming workImpact assessed; work stopped or reports recalled where neededNonconformity register
7.6 Data and information managementLIS validated, access controlled, data protectedLIS validation, user access list, backup tests
7.7 ComplaintsComplaints investigated and closed with feedbackComplaint log
7.8 Continuity and emergency preparednessPlans for power, IT or analyser failure, testedContinuity plan, drill records

For practical guidance on 7.3.7, see our guides on internal quality control in medical labs and EQA vs proficiency testing.

Clause 8: Management system requirements

Under Option A the lab meets clauses 8.2 to 8.9 directly. Under Option B a lab with a management system established to ISO 9001, capable of supporting clauses 4 to 7, is considered to meet at least the intent of 8.2 to 8.9.

ClauseWhat auditors look forEvidence to have ready
8.2 and 8.3 Documentation and document controlCurrent, approved documents at point of useMaster list, approval and withdrawal records
8.4 Control of recordsRecords legible, protected and retained for defined periodsRetention schedule, archive
8.5 Risks and opportunitiesActions planned and evaluated for effectivenessRisk register, action plans
8.6 ImprovementFeedback sought from users, patients and staffSurvey results, improvement log
8.7 Nonconformities and corrective actionRoot cause found; effectiveness checkedCorrective action records
8.8 EvaluationsQuality indicators monitored; internal audits cover all requirementsAudit programme and reports, indicator trends
8.9 Management reviewsPlanned reviews with required inputs and outputsMinutes, action tracker

Worked example: a hospital lab in Kochi preparing for NABL

A 300-bed hospital laboratory in Kochi covers biochemistry, haematology and microbiology, plus glucometers on the wards and a blood gas analyser in ICU. Before applying for NABL accreditation to ISO 15189:2022, the quality manager runs this checklist as a gap analysis.

The gaps are typical: no impartiality risk assessment. Ward glucometers sit outside the quality system, with no operator training records. Measurement uncertainty was calculated when the main analyser was installed and never reviewed, and a recent LIS upgrade went live without checking that results transfer correctly.

Over the following months, a POCT committee is formed, operators are trained and authorised, and glucometer IQC is logged centrally. MU is recalculated from recent IQC data, and a sample of reports is compared between analyser and LIS before the change is signed off. A full internal audit and management review follow, so assessors see working evidence rather than new paperwork.

Common mistakes: the top 10 audit findings

  1. Competence assessments missing or overdue for authorised staff.
  2. A new analyser in use before method verification was complete.
  3. IQC rules applied by habit rather than chosen for each examination's performance.
  4. EQA failures closed without root cause or a review of patient results.
  5. Measurement uncertainty estimated once and never reviewed.
  6. Reference intervals copied from kit inserts without checking they suit the local population.
  7. Critical results communicated but not recorded, or outside the defined time.
  8. Calibration certificates filed without review against acceptance criteria.
  9. Generic risk assessments that ignore POCT and the lab's real processes.
  10. Corrective actions with no effectiveness check, and internal audits that skip clauses.

Key takeaway: Assessors judge ISO 15189:2022 on evidence. For every clause, know who is responsible, which record proves it, and where to find it in under a minute.

Frequently Asked Questions

What is the latest version of ISO 15189?

The current edition is ISO 15189:2022, Medical laboratories, Requirements for quality and competence, published in December 2022. It replaced ISO 15189:2012. The three-year transition period ended in December 2025, so accredited labs are expected to be assessed against the 2022 edition. Confirm your own status and any outstanding actions with your accreditation body.

What are the main changes in ISO 15189:2022?

The 2022 edition follows the structure of ISO/IEC 17025:2017, with management system requirements in clause 8. It places stronger emphasis on risk management and on patients, including a new clause on requirements regarding patients. It also incorporates point-of-care testing requirements previously found in ISO 22870, and is generally less prescriptive about named procedures.

Does ISO 15189:2022 cover point-of-care testing?

Yes. POCT requirements formerly in ISO 22870 are now included, with additional requirements in Annex A. Where your lab is responsible for POCT, expect questions on governance, operator training and authorisation, quality control and result recording. Check with your accreditation body how POCT should appear in your scope.

Is ISO 15189 accreditation mandatory?

Accreditation is generally voluntary, but some health authorities, insurers, or government schemes require or favour accredited laboratories. Requirements differ between India and the GCC states and can change, so check with your regulator and accreditation body, such as NABL in India, before planning on the basis that accreditation is optional.

What is the difference between ISO 15189 and ISO/IEC 17025?

ISO 15189 is written for medical laboratories and covers the full patient pathway, including pre-examination, clinical advice, reference intervals and patient-related requirements. ISO/IEC 17025 applies to testing and calibration laboratories in general. The 2022 edition of ISO 15189 shares much of its structure with ISO/IEC 17025:2017, which helps labs that operate under both.

Next steps

If you are preparing for assessment or leading a transition project, our ISO 15189 Advanced: Medical Laboratory Quality & Competence course works through these requirements with practical examples. For related reading, see internal quality control in medical labs, EQA vs proficiency testing and our pre-analytical errors checklist. If you are unsure where to start, speak to a counsellor.